Title: Validation Of High Accuracy Weighing Equipment
Author: Walter E. Kupper
Source: 1991 Measurement Science Conference
Year Published: 1991
Abstract: Balances and SCSlSS are tested by placing calibrated mass standards on them and ascertaining that the displayed weight values are within given tolerances. A problem arises when the equipment under test itself is the most accurate in its class, i.e., of the type that is being used to calibrate the mass standards in the first place. With a working knowledge of the principles of both the equipment and the process of weighing, it is possible to design test procedures which are simple to use yet convincingly rigorous. This paper discusses the working principles of high accuracy weighing equipment, how they manifest themselves in its performance, and what to look for in B validation test.