Title: Usm GT400Customer Presentationv3 21Mar14 7/16/14 1 A Six (6) Path Ultrasonic Gas Flowmeter With 3D-Fidelity
Author(s): [Not Available]
Source: 2014 North American Custody Transfer Measurement Conference
Year Published: 2014
Abstract: evere disturbance testing (HMP) did not pass within the OMIL limits AND without a flow conditioner. Class 1 all other testing. NO Flow Conditioner. The recorded datasets have shown that there was a difference to be distinguished, with on-line real-time diagnostics, between the high level disturbance at 10D inlet length compared to the same disturbance at a 20D inlet. The 10D datasets that were recorded were outwith the OIML Class 1 limits there were deviations especially at high flow rates that were more than 0.5% however the general behavior deviated significantly from the baseline. After 20D the effect of the disturbance is reduced and only one of the three datasets had two values at low flow rates that were outwith the Class 1 limit by less than 0.05%