Title: Inter-Laboratory Comparisons Between Nmij And Accredited Laboratories For Water Flow In Japan
Author: N. Furuichi K.H. Cheong Y. Terao
Source: Flomeko 2013
Year Published: 2013
Abstract: Inter-laboratory comparisons between NMIJ and four JCSS accredited laboratories in Japan for water flow are carried out for wide flowrate range from 0.005 m3/h to 5000 m3/h using five type transfer flowmeters. The deviations of test results between NMIJ and the accredited labs are within approximately 0.1% and the averages of the deviations for each accredited labs are within 0.025%. These results indicate a high capability of the calibration facilities in the accredited labs. To investigate the reason of the deviations in the comparisons, influences of temperature, density and measurement time are observed