Email Document Reference

Enter your email address below and the reference for this document will be sent to shortly from webmaster@ceesi.com.

Title: The Mikes Measuring System For Gas Mass Flow
Author: Sampo Sillanpaa, Martti Heinonen
Source: Flomeko 2003
Year Published: 2003
Abstract: Due to the increasing need for traceable gas flow measurements in Finland, the development of a national measurement standard for gas mass flow was initiated at the Centre for Metrology and Accreditation (MIKES) in 2002. A primary calibration system based on dynamic weighing was constructed to provide traceability directly to the national mass and time standards. The MIKES gas flow measuring system includes also a commercial calibrator based on laminar flow elements. It is used as a working standard when calibrating mass flow meters for customers. The calibration range encompasses 0,4 mg/s to 625 mg/s. The relative standard uncertainty of the gravimetric system is between 0,2 % and 0,4 %.




In order to prevent spam and automated file downloads for documents within the Measurement Library, please follow the instructions below and then you will be able to email a reference to this article.





Copyright © 2025