Title: Development Of The National Standard Facility For Hydrocarbon Flow At 0.1 To 15m3/h
Author: Ryouji Doihara, Takashi Shimada, Yoshiya Terao, Masaki Takamoto
Source: Flomeko 2007
Year Published: 2007
Abstract: The NMIJ has constructed a new calibration facility in order to establish the national standard for hydrocarbon flow at 0.1 to 15 m3/h. This middle-size oil flow primary standard facility applies a static and gravimetric method with flying start and finish. A rotating double wing diverter (RDWD) and an improved weighing tank system have been developed for this new facility. After adjustment of the start and stop signal timings, the diverter timing error was certain to fall within 1 ms.




In order to prevent spam and automated file downloads for documents within the Measurement Library, please follow the instructions below and then you will be able to email a reference to this article.