Title: Internal Comparison Between Mexicos Primary Standard For Water Flow Measurements And A Compact Prover Using A Turbine Meter Transfer Standard
Author: Jose Lara Manriquez, Emmanuel Rios Carrizalez And Dario A. Loza Guerrero
Source: 2006 International Symposium on Fluid Flow Measurement
Year Published: 2006
Abstract: This paper describes the results of an internal comparison between two calibration techniques - Mexicos primary standard for liquid flow measurements, based on the static weighing principle and a compact piston prover, volumetric method - using a turbine meter of 100 mm as transfer standard. The transfer standard was calibrated at several flow rates simultaneously for both the gravimetric and the volumetric methods. It is concluded from the results that the maximum difference between these two techniques was 0,05 % at 900 L/min.