Title: Flomeko 2016, Sydney, Australia, September 26 - 29, 2016 Page 1 Research On Calibration Device Using Plunger Metering Cylinder For Turbine Flowmeter
Author: B. Wang L.Y. Wang Q.W. Cao
Source: Flomeko 2016
Year Published: 2016
Abstract: Accurate fuel rate acquisition is one of the basic requirements to improve the reliability and confidence of testing results for aero - engines. Because of its high precision and good repeatability, turbine flowmeter is widely used in flow measurement of fuel or fuel system. However, due to the influence of wear and corrosion, the meter coefficient is bound to change. To accurately evaluate the current meter coefficient of turb ine flowmeter, we developed a calibration device using a plunger metering cylinder for turbine flowmeter. Considering the mathematical model of turbine flowmeter, the stability of fuel pushing was researched through system modelling and simulation. As a re sult, the consistency between the calculated flow and the output flow rate of the calibrated meter was proved, which indicates the effectiveness of the device




In order to prevent spam and automated file downloads for documents within the Measurement Library, please follow the instructions below and then you will be able to email a reference to this article.