Author: K-H. Cheong N. Furuichi R. Doihara T. Shimada Y. Terao
Source: Flomeko 2013
Year Published: 2013
Abstract: Continuing efforts to expand the calibration flow range of small liquid hydrocarbon flow facility at NMIJ is being carried out. To justify the calibration capability of the facility in the flow range of 1 L/h 10 L/h, an in-house comparison with small water flow facility at NMIJ was conducted. Two Coriolis flowmeters were used as transfer meters. Effect of zero instability of the transfer meters on calibration results in flow range near zero was found significant and compensation of zero point flowrate was applied to reduce the effect. Sufficient number of measurements were done to reduce the uncertainty due to the reproducibility of transfer meters