Email Document Reference

Enter your email address below and the reference for this document will be sent to shortly from webmaster@ceesi.com.

Title: Chromatograph Maintenance And Troubleshooting
Author: Fred Ryel
Source: 2013 International School of Hydrocarbon Measurement
Year Published: 2013
Abstract: The GC System will perform accurately for long periods with very little attention. However a bi-monthly record of certain parameters will assist greatly in assuring that your Analyzer is operating to specifications. The GC System Maintenance Checklist should be filled in bi-monthly, dated, and kept on file for access by maintenance technicians as necessary. See next page. This gives you a historical record of the operation of your Analyzer, enables a maintenance technician to schedule replacement of gas cylinders at a convenient time, and allows quick troubleshooting and repair when it becomes necessary. A chromatogram, a configuration report, and a raw data report should also be made and filed with the checklist, furnishing a positive dated record of the Analyzer. The chromatogram and reports will also prove valuable in comparison with the chromatograms and reports run during troubleshooting