Title: Comparison results for gas flow rate between Measurement Canada and CENAM
Author: C. Lachance R Arias
Source: 2018 International Symposium on Fluid Flow Measurement
Year Published: 2018
Abstract: Results of a gas flow rate bilateral comparison between Measurement Canada (MC) and CENAM are discussed. A rotary meter, G40 was calibrated from 1.4 m3/h up to 65 m3/h both laboratories used bell provers to calibrate the meter. An average dij/% 0.12 along the calibration curve was determined. Measurement Canada initiated a research to determine the origin of the measurement differences. MC implemented a change in the bell prover calibration procedure results of the findings are documented in this paper.