Title: Inter-Laboratory Comparison Of Small Water Flow Calibration Facilities Between Ehj And Nmij
Author: N. Furuichi Y. Terao T. Shimada S. Ogawa L. Cordova
Source: Flomeko 2016
Year Published: 2016
Abstract: Comparison between facilities at E ndress+Hauser Japan and NMIJ was carried out according to three transfer standards, using a Coriolis flowmeter. The range of flowrate was from 30 k g/h to 36000 kg /h. Comparis on results showed excellent agreement for both standard s, at less than 0.018%. En values are also less than 0.52 for all examined flowrate point s. These results demonstrate the consiste ncy of facilities at both labs