Title: Kind Of New Calibrationmethod For The Volume Of Bell Prover
Author: Cui Lishui, Lixu
Source: Flomeko 2010
Year Published: 2010
Abstract: new 1m3 bell prover is developed by flow laboratory in china national institute of metrology. Laser tracker and laser interferometer is applied to measure the volume of the bell prover. The result suggests this kind of geometric method is feasible and the uncertainty of result is less than 0.1%.